dionex uhplc - bruker amazon x ion trap ms
bruker amazon x ion trap
- mass range: 50-3000, 200-6000 (extended mass range)
- positive/negative ion operation
- zero delay alternatingtm: polarity switching at 20 hz ms acquisition
- esi and apci source.
- pneumatic off-axis nebulization for flow rates up to 1ml/min
- high sensitivity rf ion guide
- manual msn up to ms11 in all scan modes.
- multiple reaction monitoring (mrm) with ms/ms and ms3
- full scan sensitivity in ms/ms: reserpine 250fg s/n>500: 1 (2ul at 125fg/ul)
- resolution fwmh (u):
- 0.58 (xtremescan)
- 0.5 (ultrascan)
- 0.30 (enhanced resolution mode)
- 0.1* (maximum resolution mode) *achieved for multiply charged ions (z>2), shown on the mono isotopic masses of the pgrp-2 peptide (3+, 4+ and 5+).
- 3.0 (extended mass range)
- scan speed (u/sec):
- 52000 (xtremescan)
- 32000 (ultrascan)
- 8100 (enhanced resolution mode)
- 4600 (maximum resolution mode)
- 27000 (extended mass range)
- 4 stage vacuum system, 24m3/h rough pump
- low pulsation syringe pump for direct injection.
dionex ultimate 3000 uhplc
dgp-3600rs serial dual-piston pump
- +/- 0.1% flow rate accuracy
- 2 pumps, 3 solvents each
- 100-5000ul recommended flow range
tcc-3000rs column compartment
- 5c to 110c temperature range
- +/-0.5℃ temperature accuracy
- +/-0.1℃ temperature stability
- +/-0.1℃ temperature precision
- 2 position, 6 port column switching valve
wps-3000trs autosampler
- inline split-loop injection
- 4-45℃ sample thermostatting
- 1ul minimum injection volume
srd-3600 6 channel degaser
- teflon af degasser membranes
- 12ml/min max flow per channel