gc-ms: varian 450gc - 300 ms/ms
- scan modes: q1ms, q3ms, precursor (parent), product(daughter), neutral loss, selected ion monitoring (sim), selected reaction monitoring (srm), full scan ionization modes: electron ionization (ei) or positive/negative chemical ionization (ci)
- source: ei/ci source with vacuum interlock for quick change of ion volumes; entrance hexapole isolates source contaminants from mass filter
- source temperature: independent control, 20 c to 325 c
- filament: single, 50 to 500 μa
- electron energy: 20 ev, 70 ev or 150 ev
- mass filters: quadruple with pre- and post-filters; lensless design maximizes transmission efficiency
- collision cell: 180 curved path with pre- and post filter region; long path ensures high dissociation efficiency
- mass range: up to 800 da
- scan rate: up to 6000 da/sec
- resolution: 0.7 da at 1250 da/sec or 0.6 da at 500 da/sec
- ei dynamic range: >10^5 (compound dependent)
- mass axis stability: 0.1 da over 24 hours
- transfer line temperature: independent control; 50 c to 325 c
- manifold temperature: independent control; 30 c to 50 c
- ion detector: positive or negative ions, constant 5 kv post-acceleration voltage
- carrier gas flow: up to 8 ml
varian 450 gc
column
factorfour capillary column: vf-5ms, 30m x 0.25mm id, df=0.25um
injectors
cp-1177 split/splitless injector
- temperature range: 50 c to 450 c isothermal
1079 ptv injector
- temperature range: -99 c to 450 c with ln2; -65 c to 450 c with lco2
- maximum temperature ramp rate: 200 c min
- temperature ramps: 4
injector efc
- pressure range: 0 to 100 psi
- total flow: 500 ml/min at 10 psi
- quickswitch valve: instantly switch between injector/ columns
- chromatoprobe: probe device for solids, large volumes of liquids, or slurries