ion chromatograph (ic)

the dionex ics-1100 is a dual channel (anions and cations) ic and includes:

  • dionex ics-1100 ion chromatograph with anion and cation self-regenerating suppressors (4mm)
  • as40 automated sampler
  • as14 column (4 x 250mm)
  • cs12a column (4 x 250mm)
  • as-dv automated sampler
  • chromeleon 7.0 - workstation

theory of operation

anions:

a sample is introduced into a stream of carbonate-bicarbonate eluent and passed via high pressure through a highly basic ion exchange column where anions are separated based on their relative affinity to the column.

cations:

a sample is introduced into a stream of 22mn h2so4 eluent and passed via high pressure through an ion exchange column where cations are separated based on their relative affinity to the column.

analysis:

the peaks with eluent are passed through self-regenerating micromembrane suppressors (anions or cations) where baseline effects from the eluent are reduced. the conductivity of individual peaks is then measured with a conductivity cell.

available methods:

current methods are for trace analysis in water.:

  • wica (anions-trace-water)
  • wica1 (anions-higher level)
  • wicc (cations-trace-water)
wica- water-anions by ic
anion (cl)range (ppm)
phosphate (o-po4)0.050...1.00
chloride (cl)0.050...1.00
nitrite (no2-n)0.025...1.0
nitrite (no2-n)0.025...1.0
nitrate (no3-n)0.025...3.2
sulphate (so4)0.250...25.0

criteria for acceptance of samples are:

  • samples do not have visible suspended solids and are not cloudy
  • maximum analytical range < 100 ppm for any analyte.
  • no organic materials, i.e. humic acid. colored samples may require cleanup before analysis.
  • sample ph between 0 and 14

acceptable sample bottles are glass or polyethylene. store at 4 degc. holding times recommended by the epa for trace anions is immediate to 48 hours (sulphate=3 days). samples must not be frozen.

wicc : water-cations by ic
anion (cl)range (ppm)
sodium (na)0.100 - 10.0
ammonium (nh4-n)0.010 - 2.50
potassium (k)0.100 - 10.0
magnesium (mg)0.100 - 10.0
calcium (ca)0.100 - 10.0

criteria for acceptance of samples are:

  • samples do not have visible suspended solids and are not cloudy
  • maximum analytical range < 100 ppm for any analyte.
  • no organic materials, i.e. humic acid. colored samples may require cleanup before analysis.
  • sample ph between 0 and 14

acceptable sample bottles are polyethylene. store at cool to 4 c. glass can cause na leaching.filtration of samples through 0.45 um filters and storage at 4 c will stabilize all analytes for 6 weeks, except nh4+ which is stable for 4 days.